V18

Reflectometer for biological applications (BioRef)



Instrument Data
Neutron guide NL 3b
Collimation
Monochromator
Wave length 90 Hz: 2.0 - 6.0 Å
45 Hz: 2.0 - 10.0 Å
25 Hz: 2.0 - 16.4 Å
Wave length resolution constant Δλ/λ = 1 - 5%, 7 - 11%
Scattering plane horizontal
Flux
Range of reflectivities 1 ·10-7 with a 50x80mm sample
q resolution ΔQ/Q = 1.4 - 7% and 10 - 15%
Detector 300 x 300 mm²
multiwire PSD detector
Maximum sample size
Polarized neutrons no
Guide field
Instrument options posibility of combiened NR and ATR-IR measuremnts
Sample environment rectangular flow cells (50x80mm)
round flow cells (Ø 60mm),
hydration chamber
Software