PINK

PINK@EMIL

The PINK branch is operated in the tender x-ray energy range (2.1 - 10 keV) providing access to non-resonant / resonant XES studies of transition metals ranging from Ti to Cu (Kα, Kβ lines) and Zr to Ag (Lα, Lβ lines), as well as light elements P, S, Cl, K, Ca (Kα, Kβ lines). A multilayer monochromator provides the high photon flux allowing measurements of very diluted substances. Taking advantage of a wavelength-dispersive von Hamos analyzer to record the entire fluorescent spectrum simultaneously, we can provide very attractive time-resolved XES measurements.

The PINK instrument is located at the hard (2000 eV – 10000 eV) x-ray branch of the EMIL beamline.

Selected Applications:
  • homo- and heterogeneous catalysis
  • electrochemistry
Figure 1: Layout of the PINK instrument (for details see text below).

Figure 1: Layout of the PINK instrument (for details see text below).


Methods

XRF

Remote access

not possible

Instrument data
Phone (~49 30 8062-) 12109
Beam availability 12h/d
Source CPMU17
Monochromator DCM Si(111), Si(311), Si(422); Multilayer mono
Energy range (at experiment)
  • DCM: 2200-10000 eV
  • Multilayer mono: 2300-9500eV
  • Energy resolution
  • DCM: E/ΔE ~ 10000 (Si111)
  • Multilayer mono: tbd
  • Flux
  • DCM: 1011-1012 ph/s
  • Multilayer mono: 1013-1014 ph/s
  • Polarisation horizontal
    Focus size (hor. x vert.) FWHM 700 x 30 µm²
    Temperature range 15K - 300K
    Pressure range 10 -2 – 100mbar, vacuum or He
    Detector Eiger, GreatEyes CCD
    Manipulators Yes (X-Y)
    Sample holder compatibility
    Additional equipment von Hamos spectrometer:
    energy resolution 0.5 - 1 eV
    time resolution 0.1 - 10 s
    Additional information Details beamline CPMU17

    Figure 1 illustrates the layout of the PINK instrument showing position in meters of the beamline components with respect to the center of the U17 undulator. Beamline optics: toroidal (R=2614923mm, r=197mm) mirror M1, optional double crystal monochromator DCM, cylindrical (r=907.5mm) multilayer monochromator M2. Motorized apertures: AU1, AU2, AU3, JJ. Filters: F1 - set of water cooled diamond attenuators (5µm, 10µm and 20µm), F2 – 3 sets by 6 attenuators each (C, Al and Ta). Optical beam position monitors: BPM1, BPM2 and BPM3. Quadrant I0 monitor: Q. Photon shatter: PH. Vacuum windows: W1- 12µm beryllium, W2, W3 – 10µm diamond. Sample holder: S. Atmospheric von Hamos spectrometer (VH#1, 5-10 keV): analyser crystal – Cr, detector – Det. Vacuum von Hamos spectrometer (VH#2, 2.2-5keV).