PM4

Soft X-ray dipole variable polarisation with PGM and MHz Chopper for LowDosePES

This soft x-ray bending magnet beamline is equipped with a Plane Grating Monochromator operated in collimated light (collimated PGM) and with a MHz Chopper System to create single bunch on demand for the ArTOF Low Dose PES end-station.


Station data
Temperature range 20-500 K by resistive heating; up to ~1200 / 1700 K long-term / flashing by electron bombardment; direct current heating also possible
Pressure range Base pressure within <5x10^-10 and 3x10^-9 mbar. With possibility to dose selected gases.
Weitere Details LowDosePES
Beamline data
Segment L08
Location (Pillar) 10.1
Source D81 (Dipole)
Monochromator PGM
Energy range 20 - 2000 eV
Energy resolution 6,000 at 400 eV
Flux 1e9 - 1e10 phot/s
Polarisation horizontal
Divergence horizontal 1.5 mrad
Divergence vertical 1 mrad
Focus size (hor. x vert.) down to 0.1 mm x 0.1 mm
User endstation not possible
Distance Focus/last valve 1050 mm
Height Focus/floor level 1415 mm
Beam availability 24h/d
Phone +49 30 8062 14713