• Gonthier, J.; Barrett, M.; Aguettaz, O.; Baudoin, S.; Bourgeat-Lami, E.; Demé, B.; Grimm, N.; Hauß, T.; Kiefer, K.; Lelièvre-Berna, E.; Perkins, A.; Wallacher, D.: BerILL: The ultimate humidity chamber for neutron scattering. Journal of Neutron Research 21 (2019), p. 65-76

10.3233/JNR-190109
Open Access Version (externer Anbieter)

Abstract:
We present the design and the performances of a precision humidity chamber for neutron diffraction which is the result of a very fruitful collaboration between the Helmholtz-Zentrum Berlin and the Institut Laue-Langevin. The sample temperature can be set between 10 and 85°C. The relative humidity is controlled from 10 to 99.9%±0.01% by controlling the temperatures of the sample and a water bath at the millidegree Celsius scale. Sample exchanges are extremely easy and the operation only requires entering two parameters: sample temperature and relative humidity. The performances are confirmed by neutron diffraction measurements performed on DMPC and DOPC.